TB 9-5855-1891-40
Figure 7. Boresight collimator mass simulator installation.
(16) Look through EYEPIECE and ensure STATIONARY RETICLE REFLECTED
IMAGE (fig. 6) is visible. If STATIONARY RETICLE REFLECTED IMAGE (fig. 6) is not
visible, perform b (6) through (9) below.
(17) Rotate SMALL CENTER MICROMETER HORIZONTAL and VERTICAL
CONTROLS (fig. 3) until VERTICAL AND HORIZONTAL BIFILARS are aligned with
STATIONARY RETICLE REFLECTED IMAGES (fig. 6).
(18) Hold SMALL CENTER HORIZONTAL and VERTICAL CONTROLS and rotate
OUTER MICROMETER HORIZONTAL and VERTICAL CONTROLS until 30 is indicated
(19) Verify that HORIZONTAL and VERTICAL BIFILARS (fig. 6) remain aligned
with stationary reticle reflected images with 30 indicated in HORIZONTAL and VERTICAL
(20) Rotate SMALL CENTER MICROMETER HORIZONTAL and VERTICAL
stationary reticle reflected image can be seen without obstruction.
(21) Move MIDDLE MIRROR OPTICAL BAFFLE (fig. 8) to open position.
10