Figure 6. Reflected image adjustment.
Boresight Collimator Test Set Accuracy

Calibration Procedure For Test Set, Boresight Collimator, Ts-3784/Tas Page Navigation
  4    5    6    7    8  9  10    11    12    13    14  
img
. . . . . . . . . . .
TB 9-5855-1891-40
Figure 7. Boresight collimator mass simulator installation.
(16) Look through EYEPIECE and ensure STATIONARY RETICLE REFLECTED
IMAGE (fig. 6) is visible. If STATIONARY RETICLE REFLECTED IMAGE (fig. 6) is not
visible, perform b (6) through (9) below.
(17) Rotate SMALL CENTER MICROMETER HORIZONTAL and VERTICAL
CONTROLS (fig. 3) until VERTICAL AND HORIZONTAL BIFILARS are aligned with
STATIONARY RETICLE REFLECTED IMAGES (fig. 6).
(18) Hold SMALL CENTER HORIZONTAL and VERTICAL CONTROLS and rotate
OUTER MICROMETER HORIZONTAL and VERTICAL CONTROLS until 30 is indicated
in HORIZONTAL and VERTICAL VIEWERS (fig. 3).
(19) Verify that HORIZONTAL and VERTICAL BIFILARS (fig. 6) remain aligned
with stationary reticle reflected images with 30 indicated in HORIZONTAL and VERTICAL
VIEWERS (fig. 3).
(20) Rotate SMALL CENTER MICROMETER HORIZONTAL and VERTICAL
CONTROLS (fig. 3) to move HORIZONTAL and VERTICAL BIFILARS (fig. 6) so that
stationary reticle reflected image can be seen without obstruction.
(21) Move MIDDLE MIRROR OPTICAL BAFFLE (fig. 8) to open position.
10