TB 9-5895-1913-40
Table 1. Calibration Description Continued.
Test instrument parameters
Performance specifications
RF output
Range: -130 dBm to +24 dBm: 10 Hz to 4 GHz
Accuracy:
≥-37 dBm to +24 dBm
10 Hz to 100 kHz
0.05 dB
10 Hz to 100 kHz
-48 dBm to <-37 dBm
0.10 dB
100 kHz to 125 MHz
-48 dBm to +24 dBm
0.20 dB
>-74 dBm to ≤-48 dBm
10 Hz to 125 MHz
0.20 dB
10 Hz to 125 MHz
>-94 dBm to -74 dBm
0.50 dB
10 Hz to 125 MHz
-130 dBm to -94 dBm
1.5 dB
>125 MHz to 300 MHz -48 dBm to +20 dBm
0.20 dB
>125 MHz to 300 MHz >-74 dBm to ≤-48 dBm
0.20 dB
>125 MHz to 300 MHz >-94 dBm to -74 dBm
0.50 dB
>125 MHz to 300 MHz -130 dBm to -94 dBm
1.5 dB
>300 MHz to 1.4 GHz -17 dBm to +20 dBm
0.25 dB
>300 MHz to 1.4 GHz >-74 dBm to ≤-17 dBm
0.50 dB
>300 MHz to 1.4 GHz >-94 dBm to -74 dBm
1.0 dB
>300 MHz to 1.4 GHz -130 dBm to -94 dBm
1.5 dB
>1.4 GHz to 3 GHz
-17dBm to +14 dBm
0.30dB
>-74 dBm to ≤-17 dBm
>1.4 GHz to 3 GHz
0.50dB
>1.4 GHz to 3 GHz
>-94 dBm to -74 dBm
1.0 dB
>1.4 GHz to 3 GHz
-130 dBm to -94 dBm
1.5 dB
>3 GHz to 4 GHz
>-74 dBm to +14 dBm
0.50 dB
>3 GHz to 4 GHz
>-84 dBm to -74 dBm
1.0 dB
1
SSB Phase Noise
Offset: 10 Hz to 10 MHz
Accuracy (dBc/Hz): -62 to -155
Not verified in this procedure.
1
EQUIPMENT REQUIREMENTS
4. Equipment Required. Table 2 identifies the specific equipment to be used in this
calibration procedure. This equipment is issued with Secondary Reference Calibration
Standards Set, NSN 4931-00-621-7878. Alternate items may be used by the calibrating
activity. The items selected must be verified to perform satisfactorily prior to use and must
bear evidence of current calibration. The equipment must meet or exceed the minimum use
specifications listed in table 2. The accuracies listed in table 2 provide a four-to-one ratio
between the standard and TI. Where the four-to-one ratio cannot be met, the actual
accuracy of the equipment selected is shown in parentheses.
5. Accessories Required. The accessories required for this calibration are common
usage accessories, issued as indicated in paragraph 4 above, and are not listed in this
calibration procedure. The following peculiar accessory is required: Termination, Agilent,
Model 11048C, verified to 50 Ω 0.2% (49.90 Ω to 50.10 Ω) at DC.