TB 9-5915-214-35
b. Connect TI to a 115 V ac source.
c. Press POWER pushbutton to ON and allow at least 20 minutes for equipment to
warm up.
15. Power Supply And dc
Adjustments
b. Adjust R105 (fig. 1), located to the left of the
rear panel CH 1 INPUT BNC
connector for digital multimeter indication of -15 V 0.05 V.
d.
Adjust R102 (fig. 1) for digital multimeter indication of
+15 V
0.05 V.
Verify digital multimeter indicates +5 V 0.2 V for both test points.
f.
NOTE
then repeat procedures using CH 2. When performing CH 2
adjustments, add the number 200 to all capacitor or resistor
references. For example referenced in step j below for CH 1 is
C167. When performing the same procedure for CH 2, the
capacitor would be C367.
a.
Press MODE parameter key until
CUTOFF FREQUENCY display indicates L.P.
b. Press INPUT(db) ∇ and OUTPUT GAIN(db) ∇ control keys until displays indicate 00.
c.
Press TYPE parameter key until
CUTOFF FREQUENCY display indicates bu.
d. Press SHIFT parameter key.
e. Press TYPE parameter key; dC should be displayed in the CUTOFF
FREQUENCY display. If
CUTOFF FREQUENCY display does not
indicate dC,
repeat
steps d and e until dC is displayed.
f. Press corresponding keys as listed in (1) through (4) below and verify CUTOFF
(1)
2 data entry key.
(2)
5 data entry key.
(3)
KILO parameter key.
(4)
FREQ parameter key.