TB 9-5915-214-35
x. Repeat p through t above.
y. Repeat technique of a through x above for CH 2.
a. Connect calibrator to INPUT CH 1 using 50 Ω termination.
b. Connect true rms voltmeter to OUTPUT CH 1.
c. Press MODE parameter key until CUTOFF FREQUENCY display indicates bYP.
d. Press corresponding keys as listed in (1) through (4) below and verify CUTOFF
FREQUENCY indicates 25.00.
(1)
2 data entry key.
(2)
5 data entry key.
(3)
KILO parameter key.
(4)
FREQ parameter key.
e. Adjust calibrator for a 200 Hz, 1 V output and reference true rms voltmeter to 0 dB.
f.
Press MODE parameter key until
CUTOFF FREQUENCY display indicates L.P.
If true rms voltmeter does not indicate 0 dB, 0.01 dB, adjust R250 (fig. 2) until true rms
g. Press corresponding keys as listed in (1) through (4) below and verify CUTOFF
FREQUENCY indicates 256.0.
(1)
2 data entry key.
(2)
5 data entry key.
(3)
6 data entry key.
(4)
FREQ parameter key.
h. Press MODE parameter key until CUTOFF FREQUENCY display indicates bYP.
i. Adjust calibrator for a 10 kHz, 1 V output and reference true rms voltmeter to 0 dB.
j.
Press MODE parameter key until
CUTOFF FREQUENCY display indicates h.P.
If true rms voltmeter does not indicate 0 dB, 0.01 dB, adjust R251 (fig. 2) until true rms
k. Repeat technique of a through j above for CH 2.
18. Output Frequency
Response
a
Press MODE parameter key until CUTOFF FREQUENCY display indicates h.P.
b. Press corresponding keys as listed in (1) through (3) below and verify CUTOFF
FREQUENCY indicates 2.000.
(1) 2 data entry key.
(2) KILO parameter key.
(3) FREQ parameter key.
c. Connect calibrator WIDEBAND output to INPUT CH 1 using cable and
termination supplied with calibrator and adjust calibrator for a 20 kHz, 0.1 V output.