• Home
  • Download PDF
  • Order CD-ROM
  • Order in Print
Harmonic Rejection
Table 4. Harmonic Rejection - Continued

Calibration Procedure For High Power Tunable Band Pass Filter Rodale Electronics, Inc. Model 13589299 Page Navigation
  1    2    3    4  5  6    7  
img
. . . . .
TB 9-5915-215-40
(5) Press the spectrum analyzer Peak Search, Marker, Delta keys.
(6) Set signal generator output frequency and spectrum analyzer center frequency
for that of the 2d harmonic as listed in table 4 and press Peak Search key. Spectrum
analyzer will indicate as listed in table 4.
(7) Repeat technique of (6) above for the 3d and 4th harmonics.
Figure 2. Harmonic rejection.
(8) Repeat technique of (2) through (7) above for the remaining thumb wheel and
frequency setting listed in table 4.
Table 4. Harmonic Rejection
Test instrument
Signal
Spectrum analyzer
generator
Center
Thumb
Harmonic
Frequency
frequency
Level indication
wheel
#
(Hz)
(Hz)
(dB)
01
---------------
135
k
135
k
Ref
50
---------------
2
270
k
270
k
50
---------------
3
405
k
405
k
50
---------------
4
540
k
540
k
02
---------------
300
k
300
k
Ref
50
---------------
2
600
k
600
k
50
---------------
3
900
k
900
k
50
---------------
4
1.2 M
1.2 M
03
---------------
600
k
600
k
Ref
50
---------------
2
1.2 M
1.2 M
50
---------------
3
1.8 M
1.8 M
50
---------------
4
2.4 M
2.4 M
04
---------------
1.0 M
1.0 M
Ref
6


Privacy Statement