TB 9-6130-285-24
Table 1. Calibration Description Continued
Test instrument parameters and performance specifications
E3645A Agilent
Range/Accuracy
Load Regulation,
Line Regulation,
Ripple and Noise
Low Range
(% of output + offset)
(% of output + offset)
Normal mode voltage
0 to +35 V/ 0 to 2.2 A dc
Voltage <0.01% + 3 mV
Voltage <0.01% + 3 mV
<1 mV rms and 8 mVp-p
High Range
Current <0.01% +
Current <0.01% +
Normal mode current
0 to 60 V/ 0 to 1.3 A dc
250 uA
250 uA
<4 mA rms
20 Hz to 20 MHz
Programming Accuracy
(% of output + offset)
Voltage <0.05% + 10 mV
Current <0.2% + 10 mA
Readback Meter Accuracy
Voltage <0.05% + 5 mV
Current <0.15% + 5 mA
E3646A HP
Range/Accuracy
Line Regulation,
Load Regulation,
Ripple and Noise
Low Range
Normal mode Voltage
(% of output + offset)
(% of output + offset)
0 to +8 V/ 0 to 3 A
<0.5 mVrms and 5 mVp-p
Voltage <0.01% + 3 mV
Voltage <0.01% + 3 mV
High Range
Current <0.01% +
Current <0.01% +
Current <4 mA rms
0 to +20 V/ 0 to 1.5 A
250 uA
250 uA
20 Hz to 20 MHz
Programming Accuracy
(% of output + offset)
Voltage <0.05% + 10 mV
(<0.1% + 25mV for output 2)
Current <0.2% + 10 mA
Readback Accuracy
(% of output + offset)
Voltage <0.05% + 5 mV
(<0.1% + 25mV for output 2)
Current <0.15% + 5 mA
(<0.15% + 10 mA for output
2)
1 Maximum current limited to 5 A maximum current if using the decade resistor.
EQUIPMENT REQUIREMENTS
4. Equipment Required. Table 2 identifies the specific equipment to be used in this
calibration procedure. This equipment is issued with Secondary Transfer Calibration
Standards Set AN/GSM-286; AN/GSM-287; or AN/GSM-440. Alternate items may be used
by the calibrating activity. The items selected must be verified to perform satisfactorily
prior to use and must bear evidence of current calibration. The equipment must meet or
exceed the minimum use specifications listed in table 2. The accuracies listed in table 2
provide a four-to-one ratio between the standard and TI. Where the four-to-one ratio
cannot be met, the actual accuracy of the equipment selected is shown in parentheses.
5. Accessories Required. The accessories required for this calibration are common
usage accessories, issued as indicated in paragraph 4 above, and are not listed in this
calibration procedure. The following peculiar accessories are also required for this
calibration: Decade Resistor, Winslow, Model 336 (7907234) or Clarostat, Model 240C
(240C) or an electronic load EL750B or DLP130-50-2500, if one can be obtained.