Load and Line Regulation (Constant Voltage)
Figure 3. Load and line regulation (constant current).

Calibration Procedure For Dc Power Supplies (General) Page Navigation
  2    3    4    5    6  7  8    9    10    11    12  
img
. . . . . . .
TB 9-6130-285-24
(12) Adjust the autotransformer until the meter indicates 115 V ac.
(13) Using TI front panel controls, program voltage and current outputs for 0.
No adjustment can be made for 1791; for all others see
b. Adjustments.
appendix A.
DECADE RESISTOR
TI
+S
COM
OUT
-S
POTENTIAL
AUTOTRANSFORMER
100A
10A
COM
DC CURRENT
HI
SHUNT
LO
MULTIMETER NO. 1
Figure 2. Load and line regulation (constant voltage).
10. Load and Line Regulation (Constant Current)
a. Performance Check
(1) Connect equipment as shown in figure 3.
(2) Set the decade resistor resistance for maximum current at maximum voltage
from ratings in table 1.
(3) Using TI front panel controls, program voltage and current outputs for full scale
output (see table 1).
(4) Reduce the decade resistor resistance until TI front panel display indicates full
scale A and ensure CC enunciator is illuminated.
NOTE
In the following step, if TI does not remain in the CC mode,
decrease the decade resistor resistance until TI CC enunciator
remains on.
(5) Divide the multimeter indication by the DC current shunt certified resistance
and record results.