Figure 5. Ripple and noise calibration constant voltage.
Front Panel Control Calibration

Calibration Procedure For Dc Power Supplies (General) Page Navigation
  6    7    8    9    10  11  12    13    14    15    16  
img
. . . . . .
TB 9-6130-285-24
SECTION IV
CALIBRATION PROCESS USING ELECTRONIC LOAD
13. Preliminary Instructions
a. The instruction outlined in paragraphs 6 and 7 are preparatory to the calibration
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
name as listed in table 2.
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure.  Additional
maintenance information is contained in the manufacturer's manual for this TI.
d. Unless otherwise specified, all controls and control settings refer to the TI.
14. Equipment Setup
WARNING
HIGH VOLTAGE is used or exposed during the performance of
this calibration.
DEATH ON CONTACT may result if
personnel  fail  to  observe  safety  precautions.  REDUCE
OUTPUT(S)  to  minimum  after  each  step  within  the
performance check where applicable.
a. Connect TI to autotransformer. Connect autotransformer to 115 V ac source and set
to minimum.
b. Connect equipment as shown in figure 1.
Figure 1. Front panel control calibration.
12