TB 9-6625-011-24
Test instrument
Calibrator
RANGE
ERROR display
switch settings
indications
(VOLTS)
Meter indication scale
Initial output
< (%)
0-to-1
0-to-3
Voltage
Frequency
(Hz)
1.01
.001
1
---
1.0
V
400
.003
---
3
3.0
V
400
1.0
Reset calibrator and remove ac divider from setup.
.03
---
3
30
mV
400
1.0
.1
1
---
100
mV
400
1.0
1.02
.3
---
3
0.3
V
400
1.03
1
1
---
1.0
V
400
1
.8
---
0.8
V
400
1.25
1
.6
---
0.6
V
400
1.7
1
.4
---
0.4
V
400
2.5
1
.2
---
200
mV
400
5.0
3
---
3
3.0
V
400
1.0
10
1
---
10
V
400
1.0
30
---
3
30
V
400
1.0
100
1
---
100
V
400
1.0
300
---
3
300
V
400
1.0
Record this value for use in table 4 as ERRref1 (include the sign).
1
Record this value for use in table 4 as ERRref2 (include the sign).
2
3Record this value for use in table 4 as ERRref3 (include the sign).
(8) Connect TI INPUT to calibrator WIDEBAND output and press calibrator
WBND pushbutton.
(9) Set TI RANGE switch to .001 VOLTS.
(10) Set calibrator output for an initial 0.001V @ 400 Hz. Adjust the calibrator
output amplitude for precisely `1' on the TI meter. Press calibrator NEW REF pushbutton.
(11) Set calibrator frequency to 20 Hz, then readjust the calibrator output amplitude
for precisely `1' on the TI meter. Record calibrator Error display indication as ERRmeas1
(include the sign). Algebraically calculate TI frequency response error using the formula
below. The TI error will be < 5.0 percent.
TI frequency response error = ERRmeas1 + ERRref1
Example:
ERRmeas1 = -2.3%
ERRref1 = 0.4%
-2.3+0.4=-1.9%
(12) Repeat technique of (11) above for remaining frequencies listed for the .001
specified limits; if not, perform b (9) through (11) below.
(13) Repeat technique of (9) through (12) above for TI 1 VOLTS RANGE switch
settings and calibrator initial voltage listed in table 4. Substitute ERRmeas3 and ERRref3 in
formula. TI frequency response error will be within the limits specified; if not, perform b
(12) through (14).