TB 9-6625-011-24
(14) Vary calibrator frequency between 100 kHz and 10 MHz and, if required,
readjust C102 (fig. 1) for best in-tolerance condition.
(15) Set TI RANGE switch to .3 VOLTS. Set calibrator wideband output to 400 Hz
and adjust amplitude for a 0.3 Volt reference indication. Press NEW REF on calibrator.
(16) Set calibrator frequency to 3.0 MHz. Adjust C303 (fig. 1) for the 400 Hz 0.3 V
reference established in b (15) above (R).
(17) Vary calibrator frequency between 3 and 10 MHz and, if required, readjust
C303 (fig. 1) for best in-tolerance condition.
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.