TB 9-6625-096-50
(c) AUDIO SELECTOR switch to 1,000 ∼.
NOTE
Frequency counter (A3) must be slightly tuned off zero beat to
(3) Adjust oscilloscope controls as necessary. Oscilloscope display will be as shown
Figure 5. Test instrument - 30-percent modulation waveform.
(4) Turn AUDIO SELECTOR switch to 30 ∼,
(5) Repeat (3) above.
(6) Turn AUDIO SELECTOR switch to 9960 FM.
(7) Repeat (3) above.
NOTE
150 Hz modulation will be checked during the performance of
b. Adjustments.
Adjust LEVEL SET
adjustments (fig. 3) as
necessary to
produce
13. Tone Localizer Centering
a. Performance Check
(1) Connect AUDIO OUT connector of TI to true rms voltmeter (A8) using cable
(B8) and adapter (B1).
(2) Position controls of TI as indicated in (a) and (b) below:
(a) AUDIO SELECTOR- switch to 90 ∼.
(b) LOC-GS control fully counterclockwise.
11