TB 9-6625-102-24
IDENTIFICATION AND DESCRIPTION
1. Test Instrument Identification.
This bulletin provides instructions for the
manual was used as the prime data sources in compiling these instructions. The
equipment being calibrated will be referred to as the TI (test instrument) throughout this
bulletin.
a. Model Variations. None.
b. Time and Technique. The time required for this calibration is approximately 2
hours, using the dc and low frequency technique.
2. Forms, Records, and Reports. Forms, records, and reports required for calibration
personnel at all levels are prescribed by TB 750-25.
3. Calibration Description. TI parameters and performance specifications which
pertain to this calibration are listed in table 1.
Table 1. Calibration Description.
Test instrument parameters
Performance specifications
DCV
Accuracy
Range
(% Reading +Counts)
50 mV
0.05 + 50
500 mV
1000 mV
0.025 + 5
5V
50 V
500 V
0.03 + 5
1000 V
Accuracy
Range
(% Reading +Counts)
500 Ω
0.05 + 10
5 kΩ
50 kΩ
0.05 + 5
500 kΩ
5 MΩ
0.15 + 5
50 MΩ
1+5
3 + 10 <200 MΩ
500 MΩ
8 + 10 >200 MΩ
500 nS
1 + 10