Table 2. Minimum Specifications of Equipment Required.
DC Voltage and Linearity Accuracy

Calibration Procedure For Multimeter Apn 13666158 And Model U1252A Page Navigation
  1    2    3    4  5  6    7    8    9    10  
img
. . . . . .
TB 9-6625-102-24
Table 2. Minimum Specifications of Equipment Required Continued.
Manufacturer and model
Common name
Minimum use specifications
(part number)
CALIBRATOR
ACI
Fluke, Model 5720A (5720A)
- Continued
Range: 500 uA @ 1 kHz Accuracy 0.19%
w/amplifier, Fluke 5725A/AR
Range: 500 uA @ 10 kHz Accuracy 0.2%
(5725A/AR) (p/o MIS-35947)
Range: 5000 uA @ 1 kHz Accuracy 0.19%
Range: 5000 uA @ 10 kHz Accuracy 0.2%
Range: 50 mA @ 1 kHz Accuracy 0.19%
Range: 50 mA @ 10 kHz Accuracy 0.2%
(0.38%)
Range: 440 mA @ 45 Hz Accuracy 0.26%
Range: 440 mA @ 1 kHz Accuracy 0.39%
Range: 5 A @ 1 kHz Accuracy 0.19%
Range: 3 A @ 5 kHz Accuracy 0.8%
Range: 10 A @ 1 kHz Accuracy 0.23%
Resistance:
Range: 10 MΩ 0.26%
CAPACITANCE
Range: 10 to 1000 nF
IET, Model HACS-Z-B-6E-1pF p/o
STANDARD NO. 1
Accuracy: 0.26%
HACS-Z-B-10E-1pF
IET, Model HACS-Z-B-3E-1μF p/o
CAPACITANCE
Range:10 uF to 10 mF
STANDARD NO. 2
Accuracy: Test Report Value
HACS-Z-B-10E-1pF
FREQUENCY
Range: 100 to 4800 Hz
Agilent, Model 53131A
COUNTER
Accuracy: 0.001%
w/opt 010(AN/USM459B)
Agilent, Model 33250A (33250A)
FUNCTION
Square Wave
GENERATOR
Range: 0.2 to 0.6 VRMS
Accuracy: NA
Frequency range: 100 kHz to 10 MHz
Accuracy: 0.013%
MULTIMETER
ACV
Frequency: 4.8 kHz
Hewlett Packard, Model 3458A
Amplitude: 2.8V RMS
(3458A)
Accuracy: 1.79%
Resistance
Range
Accuracy
5 MΩ
0.02%
50 MΩ
0.25%
500 MΩ
2.01%
Range: 500 Ω to 500 kΩ
RESISTANCE
IET Labs, HARS-LX-9-0.001
STANDARD NO. 1
Accuracy: 0.02%
(HARS-LX-9-0.001)
Range: 3 to 500 MΩ
RESISTANCE
IET Labs, HRRS-Q-4-1M-10KV-
STANDARD NO. 2
Accuracy: 0.06% Multimeter Accuracy
RM (HRRS-Q-4-1M-10KV-RM)
SECTION III
CALIBRATION PROCESS
6. Preliminary Instructions
a. The instructions outlined in paragraphs 6 and 7 are preparatory to the calibration
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
name as listed in table 2.