TB 9-6625-102-24
b. Entering Adjustment Values
NOTE
Some ranges do not allow custom values to be entered.
(1) Use the ◄RANGE and ►DUAL keys to select the primary display segment to
be edited.
(2) Use the ▲HOLD and ▼NULL keys to select the desired digits.
(3) Repeat steps (1) and (2) above for all required display segments.
(4) Press the Hz key when done to start the adjustment.
(5) CAL will flash in the secondary displayed to indicate calibration is in process.
(6) PASS will flash in the secondary display when calibration process is completed.
c. Adjustment Procedure
(1) Connect stimulus to appropriate inputs.
(2) Set rotary switch to
V function.
(3) The primary display will show the reference value of the Cal Item.
(4) Using the ▲HOLD and ▼NULL keys select the first Cal Range listed in
(5) Set the calibration standard stimulus to the corresponding value listed in the
(6) Enter the actual applied input using technique in paragraph b above when a
custom value is applied (i.e. 10.086 mF).
(7) Press the Hz key to accept stimulus.
(8) When TI displays donE, reduce standard output to minimum.
(9) Repeat technique of (1) through (8) above for each Function and Cal ranges listed
Table 15. Alignment Group 1.
TI
Calibration standard stimulus
Function
Cal Range
Cal Item
Output
V
5V
0.3000 V
0.3V, 1 kHz
3.0000 V
3 V, 1 kHz
3.0000 V
3 V, 50 kHz
50 V
03.000 V
3 V, 1 kHz
30.000 V
30 V, 1 kHz
30.000 V
30 V, 50 kHz
500 V
030.00 V
30 V, 1 kHz
300.00 V
300 V, 1 kHz
300.00 V
300 V, 50 kHz
1000 V
0030.0 V
30 V, 1 kHz
0300.0 V
300 V, 1 kHz
0300.0 V
300 V, 50 kHz
16