TB 9-6625-102-24
Table 11. Resistance High.
TI
Indicates within standard
Range (Ω)
Standard No.
value recorded in
2
5M
3.0 k
2
50 M
505 k
2
500 M
40.10 M
(11) Disconnect equipment set-up.
b. Adjustments. Perform entire calibration procedure before performing alignment
a. Performance Check
ΩTEMPVmV and COM
(1) Connect function generator output to TI
inputs using a 50 Ω termination.
(2) Set the TI as listed in (a) through (c) below:
(a) Set rotary switch to Hz
.
(b) Press SHIFT key for a Hz indication.
(c) Press the RANGE key for a -1- range (Hz indication).
(4) TI will indicate within the limits of table 12; if not, perform b below.
(5) Repeat technique of (2)(c) and (4) above for remaining signal generator voltage
Function generator
TI
Voltage (mV RMS)
Min
Max
480
1k
0.9995 kHz
1.0005 kHz
200
100 k
99.95 kHz
100.05 kHz
1
600
10 M
9.995 MHz
10.005 MHz
Press the RANGE key for a -100- range (MHz indication).
1
(6) Set function generator output to minimum and disconnect equipment set-up.
b. Adjustments. Perform entire calibration procedure before performing alignment
a. Performance Check
NOTE
To minimize lead capacitance use shortest banana leads available.
(1) Connect Capacitance Standard No. 1 output HIGH and LOW to TI
ΩTEMPVmV and COM inputs and set all capacitance standard decades to zero.