TB 9-6625-102-24
Table 15. Alignment Group 1 Continued.
TI
Function
Cal Range
Cal Item
Output
open
oPEn
open
A
500 A
300.00 A
300 A dc
5000 A
3000.0 A
3000 A dc
500 A
030.00 A
30 A 1 kHz
SHIFT to
A
300.00 A
300 A 1 kHz
5000 A
0300.0 A
300 A 1 kHz
3000.0 A
3000 A 1 kHz
open
oPEn
open
mAA
50 mA
30.000 mA
30 mA
440 mA
300.00 mA
300 mA
5 A5,6
3.0000 A
3A
10 A
10.000 A
10 A
7,8
03.000 mA
3 mA, 1 kHz
50 mA
SHIFT to
mA
30.000 mA
30 mA, 1 kHz
440 mA
030.00 mA
30 mA, 1 kHz
300.00 mA
300 mA, 1 kHz
5
0.3000 A
0.3 A 1 kHz
5A
SHIFT to
A
6
3.0000 A
3 A, 1 kHz
10 A
3.0000 A
3 A, 1 kHz
10.000 A
10 A, 1 kHz
Set the calibrator to 10 MΩ output. Calibrator displayed output meets required specifications.
1
To minimize lead capacitance use shortest banana leads available.
2
4 Apply stimulus and allow 5 minutes for temperature stabilization before pressing Hz key.
5 Move connection from A mA and COM to A and COM inputs.
6 Move connection from calibrator AUX HI and LO output to amplifier HI and LO output.
7 Move connection from A and COM to A mA and COM.
8 Move connections from amplifier HI and LO outputs to calibrator AUX HI and LO outputs.
(10) Press and hold SHIFT and Hz to store new correction factors and exit.
(11) Reduce outputs to minimum and disconnect equipment setup.
a. De-energize and disconnect all equipment.
b. If TI is a stand-alone item annotate and affix DA label/form in accordance with
TB 750-25.
c. If TI is a component of SWICE kit, follow final procedure guidance in
TB 9-4920-468-24.