TB 9-6625-103-24
IDENTIFICATION AND DESCRIPTION
1. Test Instrument Identification.
This bulletin provides instructions for the
specifications were used as the prime data source in compiling these instructions. The
equipment being calibrated will be referred to as the TI (test instrument) throughout this
bulletin.
a. Model Variations. None.
b. Time and Technique. The time required for this calibration is approximately 5
hours, using the dc and low frequency and microwave techniques.
2. Forms, Records, and Reports. Forms, records, and reports required for calibration
personnel at all levels are prescribed by TB 750-25.
3. Calibration Description. TI parameters and performance specifications which
pertain to this calibration are listed in table 1.
Table 1. Calibration Description.1
Test instrument parameters
Perform specifications
Frequency
Range:
10 MHz to 30 GHz
Accuracy: < 1 part in 10-7
Time base stability: < 1.5 part in 10-9 per day
< 1 part in 10-6 per year
Line stability:
< 0.1 ppm 10% line voltage change
Output Level
Frequency:
Level:
Accuracy:
10 MHz to 20 GHz -70 to +13 dB 1.5 dB
20 to 30 GHz
-70 to +7 dB
1.5 dB
No specification for output levels -130 to -70.
Single Sideband Phase Noise
Carrier Frequency:
≤ -100 dBc/Hz at 20 kHz offset
10 MHz to 6 GHz
≤ -95 dBc/Hz at 20 kHz offset
6 to 12 GHz
≤ -89 dBc/Hz at 20 kHz offset
12 to 24 GHz
≤ -83 dBc/Hz at 20 kHz offset
24 to 30 GHz
Harmonics:
Harmonics
< -28 dBc THD
10 MHz to ≤2 GHz
< -30 dBc THD
>2 GHz to ≤30 GHz
Sub-harmonic
< -80 dBc from -70 dBm to +6 dBm
10 MHz to ≤ 500 MHz
< -40 dBc from -70 dBm to +6 dBm
>500 MHz to ≤ 30 GHz
Non-harmonic
< -50 dBc with 5 kHz offset from Center Carrier Frequency.