TB 9-6625-103-24
Table 10. Single Sideband Phase Noise.
Measuring receiver
TI FREQ
SSB avg. indication
≤ -100 dBc
10
MHz
≤ -100 dBc
5.9 GHz
≤ -95 dBc
6.1 GHz
≤ -95 dBc
11.9 GHz
≤ -89 dBc
12.1 GHz
≤ -89 dBc
23.9 GHz
≤ -83 dBc
24.1 GHz
≤ -83 dBc
1
26
GHz
≤ -83 dBc
30
GHz
1 Upper limit if using an E4440A as the measuring
receiver system.
(5) Reduce output to minimum.
a. Performance Check
(1) Connect TI RF Out to measuring receiver RF INPUT.
(2) Set measuring receiver to measure frequency modulation at 10 MHz with PEAK
+-/2 detector.
(3) Press TI keys as listed in (a) through (f) below.
(a)
Preset.
(b)
RF On/Off to Off.
(c)
AMPD, -, 1, 0, and dBm.
(d)
FREQ, 1, 0, MHz.
(e)
RF On/Off to On.
(f)
AM, AM RATE, 4, 0, 0, Hz, AM DEPTH, 3, 0, % and AM Off/On to On.
(4) Verify that the measuring receiver indication is <10 kHz.
(5) Set measuring receiver to acquire frequency then measure amplitude
modulation, with PEAK +-/2.
(6) Verify that the measuring receiver indicates within limits listed in table 11.
(7) Repeat technique of (3), (5) and (6) above for remaining rows in table 11.
TI
Measuring receiver indication
FREQ
AM rate
AM depth (%)
Min (%)
Max (%)
10 MHz
400 Hz
30
24.00
36.00
70
64.00
76.00
1 kHz
70
61.90
78.10
30
23.10
36.90
20 GHz
400 Hz
70
64.00
76.00
30
24.00
36.00
1 kHz
30
23.10
36.90
70
61.90
78.10