TB 9-6625-103-24
a. Performance Check
Figure 3. Pulse Modulation Hook-up.
(2) Press TI keys as listed in (a) through (f) below.
(a) Preset.
(b) RF On Off to off.
(c) AMPTD, 0, and dBm.
(d) FREQ, 5, and GHz.
(e) RF On Off to on.
(f) Pulse, and Pulse Off/On to on.
(a) Vertical 1 Input 50 Ω to on (lit).
(b) Vertical 1 scale to 2 mV.
(c) Trigger Source to 1.
(d) Trigger Sweep to Auto.
(e) Trigger Slope to
.
(f) Horizontal sweep speed to 10 nS.
(4) Measure rise time and verify that the oscilloscope indication is less than the rise
(5) Select the remaining oscilloscope measurement types listed in table 19 and
repeat technique of (4) above for the appropriate measurement type.
Table 19. Pulse Characteristics.
Oscilloscope measurement type
Maximum indication
Rise Time
25 nS
Overshoot
20 %
Fall Time
25 nS
(6) Reduce output to minimum, disconnect equipment setup, and reconnect as
shown in figure 4 using the OPT 526 power sensor. Zero and cal the power meter.