TB 9-6625-109-40
EQUIPMENT REQUIREMENTS
4. Equipment Required. Table 2 identifies the specific equipment to be used in this
calibration procedure. This equipment is issued with Secondary Reference Calibration
Standards Set, NSN 4931-00-621-7878. Alternate items may be used by the calibrating
activity. The items selected must be verified to perform satisfactorily prior to use and
must bear evidence of current calibration. The equipment must meet or exceed the
minimum use specifications listed in table 2. The accuracies listed in table 2 provide a
four-to-one ratio between the standard and TI.
5. Accessories Required. The accessories listed in table 3 are issued as indicated in
paragraph 4 above and are used in this calibration procedure. The following peculiar
accessories are also required: Instrument controller and the PRTCoefficient.exe program,
which is available at the USATA home page website. When necessary, these items may be
substituted by equivalent items, unless specifically prohibited.
Table 2. Minimum Specifications of Equipment Required
Minimum use
Manufacturer and model
Common name
specifications
(part number)
Agilent, Model 3458A (3458A)
Range: (100 Ohms nominal)
Accuracy: (15ppm of reading
+6ppm of range)
Accuracy: ≤ 0.005 C
THERMOMETER READOUT
Hart Scientific, Model 1590
THERMOMETER SPRT
Range: 30 C to 400 C
Rosemont, Model 162C or 162CE
Accuracy: 0.0125 C
Accessories Temperature Sources
Manufacturer and model
Common name
Purpose
(part number)
BATH, CONSTANT TEMPERATURE
+200 C
Hart Scientific, Model 6045/6050
BATH, CONSTANT TEMPERATURE
+400 C
Hart Scientific, Model 6045/6050
LOW TEMPERATURE BATH
30 C
Hart Scientific, Model 7341
LOW TEMPERATURE BATH
+30 C
Hart Scientific, Model 7341
TEMP STD MINI TPW
Triple-Point of Water
Hart Scientific, Model 9210