TB 9-6625-1176-24
(5) Repeat technique of (2) above for settings and indications listed in table 9.
b. Adjustments. Adjust A4 R604 so that the voltage difference between TP1 and TP5
(fig. 7) is 22V 220 mV (R).
Table 9. Power Supply Check
Multimeter
Test instrument
indications (V)
Voltage
Test point
Min
Max
-5.2
Marked on PC board
4.94
5.46
+15
Marked on PC board
14.25
15.75
-15
Marked on PC board
14.25
15.75
+5
Marked on PC board
4.75
5.25
15. Final Procedure
a. Deenergize and disconnect all equipment and reinstall protective cover on TI.
b. Annotate and affix DA label/form in accordance with TB 750-25.