Equipment Setup
Table 5. Dc Current Source Accuracy.

Calibration Procedure For Digital Multimeter Model 787 Process Meter Page Navigation
  1    2    3  4  5    6    7  
img
. . . . . . .
TB 9-6625-121-24
9. Ac Voltage
a. Performance Check
and COM inputs.
(1) Connect calibrator to TI V Ω
(2) Set TI rotary switch to V~.
(3) Set calibrator for an output of 360.0 mV at a frequency of 45 Hz. If TI does not
indicate within limits specified in first row of table 4, perform b below.
(4) Repeat technique of (3) above for remaining settings and indications listed in
table 4. If TI does not indicate within limits specified in table 4, perform b below.
Table 4. Ac Voltage Accuracy.
Calibrator output
Test instrument indications
Amplitude
Frequency
Range
Min
Max
360.0 mV
45  Hz
400
mV
355.3
364.7
60  Hz
357.1
362.9
200  Hz
355.3
364.7
500  Hz
334.4
385.6
3.6
V
45  Hz
4
V
3.553
3.647
60  Hz
3.573
3.627
200  Hz
3.553
3.647
500  Hz
3.344
3.856
5
V
45  Hz
40
V
4.90
5.10
12
V
60  Hz
11.90
12.10
20
V
60  Hz
19.84
20.16
28
V
60  Hz
27.78
28.22
36
V
60  Hz
35.73
36.27
200  Hz
35.53
36.47
500  Hz
33.44
38.56
360
V
45  Hz
400
V
355.3
364.7
60  Hz
357.3
362.7
200  Hz
355.3
364.7
500  Hz
334.4
385.6
900
V
45  Hz
100
V
885
915
60  Hz
892
908
200  Hz
885
915
500  Hz
833
967
(5) Set calibrator output to minimum.
b. Adjustments. No adjustments can be made.
10. Dc Current
a. Performance Check
(1) Connect calibrator to TI mA and COM inputs.
(2) Set rotary switch to mA/A
~.
(3) Set calibrator output for 27 mA. TI will indicate between 26.984 and 27.016 mA dc.
5