Table 9. Ac Volts - Continued

Calibration Proceudre For Digital Voltmeter, Model 3440A With 3444A And 3445A Plug-Ins And Digital Multimeter, Model 360 Page Navigation
  12    13    14    15    16  17
img
. . . . . . . . .
TB 9-6625-1314-24
Table 10. Ac Current
Test instrument
Electrical test set indications (mA)
Range pushbutton
Indications (mA)
Min
Max
2
1.999
1.9790
2.0190
20
19.99
19.790
20.190
200
199.9
197.90
201.90
2000
1999
1979.0
2019.0
b. Adjustments. No adjustments can be made.
20. Resistance
a. Performance Check
pushbutton and .2 range pushbutton.
(1) Press K
(2) Connect resistance standard between TI COM and MA/K
terminals.
(3) Adjust resistance standard to obtain an indication of 199.9 on TI. If decade
resistance standard does not indicate between 198.9 and 200.9 ohms, perform b below.
(4) Repeat technique of (3) above using settings and indications listed in table 11.
If decade resistance standard does not indicate within limits perform corresponding
adjustments listed in table 11.
Table 11. Resistance Accuracy
Test instrument
Resistance standard
indication (k )
Range
Adjustment
pushbutton
Indication
Min
Max
(fig. 3)
2.009 k
---
2
1.999
1.989 k
20.09 k
20 K OHM ADJ (R)
20
19.99
19.89 k
200.9 k
---
200
199.9
198.9 k
2013 k
2 M OHM ADJ (R)
2000
1999
1985 k
b. Adjustments
(1) Adjust decade resistance standard to 199.90 ohms.
(2) Adjust .2k
(fig. 3) to obtain an indication of 199.9 on TI (R).
21. Final Procedure
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.
17/(18 Blank)