TB 9-6625-1314-24
WARNING
HIGH VOLTAGE is used or exposed during the performance of
DEATH ON CONTACT may result if
personnel fail to observe safety precautions.
REDUCE
OUTPUT(S) to minimum after each step within the
performance checks where applicable.
b. Connect autotransformer to ac power source and adjust for 115 V output.
c. Set TI POWER switch to ON and allow at least 30 minutes for warm-up.
d. Press D.C. function and .2 range switch.
e. Connect TI VOLT and COM terminals.
f. If necessary adjust ZERO control (fig 3) to obtain an indication of 000.0 (00.0 on
some models) on TI.
g. Remove lead from VOLT and COM terminals.
Figure 3. Digital multimeter - adjustment locations.