Final Procedure
Dc Volts and Stability

Calibration Proceudre For Digital Voltmeter, Model 3440A With 3444A And 3445A Plug-Ins And Digital Multimeter, Model 360 Page Navigation
  8    9    10    11    12  13  14    15    16    17  
img
. . . .
TB 9-6625-1314-24
15. Equipment Setup
WARNING
HIGH VOLTAGE is used or exposed during the performance of
this calibration.
DEATH ON CONTACT may result if
personnel fail to observe safety precautions.
REDUCE
OUTPUT(S)  to  minimum  after  each  step  within  the
performance checks where applicable.
a. Connect TI to autotransformer.
b. Connect autotransformer to ac power source and adjust for 115 V output.
c. Set TI POWER switch to ON and allow at least 30 minutes for warm-up.
d. Press D.C. function and .2 range switch.
e. Connect TI VOLT and COM terminals.
f.  If necessary adjust ZERO control (fig 3) to obtain an indication of 000.0 (00.0 on
some models) on TI.
g. Remove lead from VOLT and COM terminals.
Figure 3. Digital multimeter - adjustment locations.