TB 9-6625-1314-24
(7) Remove short from INPUT terminals, and connect calibrator (in standby mode)
to TI INPUT terminals.
(8) Set INT CHECK 8000 adjustment to its mechanical midrange (approximately
five turns from fully cw (clockwise) position).
(9)
Adjust calibrator for an output of 8.000 V.
(10) Adjust A3R51 RAMP SLOPE ADJ (fig. 1) to obtain an indication of 8.000 on TI.
Adjust INT CHECK 8000 for fine adjustment if required (R).
(11) Set RANGE switch to 100 V.
(12) Adjust calibrator for an output of 80.000 V.
(13) Adjust A1R7 100V ADJ (fig. 1) to obtain an indication of 80.00 on TI (R).
(14) Set RANGE switch to 1000 V.
(15) Adjust calibrator for an output of 800.00 V.
(16) Adjust A1R8 1000V ADJ (fig. 1) to obtain an indication of 800.00 on TI (R).
(17) Adjust calibrator for a negative output of - 8.000 V.
(18) Set RANGE switch to 10 V.
(19) Adjust INT CHECK 8000 adjustment to obtain an indication of -8.000 on TI (R).
(20) Press and hold INT CHECK 8000 pushbutton.
(21) Adjust A3R60 -8.000 REF ADJ (fig. 1) to obtain an indication of -8.000 on TI (R).
Release INT CHECK 8000 pushbutton.
NOTE
If plug in model 3445A go to paragraph 11, for 3444A continue
with next step (22).
(22) Adjust calibrator output to standby and remove. Short TI INPUT terminals.
(23) Set RANGE switch to 100 MV and adjust front panel ZERO adjust to obtain
an indication of 00.00 on TI (R).
(24) Set RANGE switch to 1000 MV and adjust A2R31 (fig. 1) to obtain an
indication of 0000 (R).
(25) Remove short installed in (22) above and connect calibrator.
(26) Adjust calibrator for an output of 800.0 mV.
(27) Adjust A2R36 (fig. 1) to obtain an indication of 800.0 (R).
(28) Set calibrator for negative output and repeat (26) above.
(29) If required, readjust A2R36 and repeat (25) through (28) above until TI
indicates between 799.9 and 800.1 for both positive and negative outputs of calibrator.
(30) Connect calibrator for positive output.
(31)
Adjust calibrator for an output of 80.00 mV.
Set RANGE switch to 100 mV.
(32)
(33)
Adjust A2R34 (fig. 1) to obtain an indication of 80.00 on TI (R).