TB 9-6625-1314-24
(4) Adjust calibrator for an output of 1.9990 V.
(5) Adjust 2 F.S. (2VDC) (fig. 3) to obtain an indication of 1.9990 on TI.
17. Dc Current
a. Performance check
Connect TI COM and MA/K
terminals to calibrator current outputs.
(1)
Press TI M.A. function pushbutton and .2 range pushbutton.
(2)
(3)  Adjust calibrator to obtain a 199.9 indication on TI. Calibrator will indicate
between 199.1 mA and 200.7 mA.
(4)  Repeat technique of (2) and (3) above using settings and indications listed in
table 8. Calibrator will indicate within limits specified in table 8.
Table 8. Dc Current
Calibrator output indication
Test instrument
Indication
Range pushbutton
(mA)
Min
Max
2
1.999
1.9910 mA
2.0070 mA
20
19.99
19.910 mA
20.070  mA
200
199.9
199.10
mA
200.70
mA
2000
1999
1.9830  A
2.0150
A
b. Adjustments. No adjustments can be made.
18. Ac Volts
a. Performance Check
(1) Connect calibrator HI and LO outputs to TI.VOLT and COM terminals.
(2) Press TI A.C. function and 2 range pushbuttons.
(3) Adjust calibrator output frequency to 100 Hz and voltage to obtain an indication
of 1.999 on TI. If calibrator output does not indicate between 1.9790 V and 2.0190 V ac,
perform b (1) and b (2) below.
(4) Repeat technique of (2) and (3) above using settings and indications listed in
table 9. Ac voltage calibrator will indicate within limits specified.
Table 9. Ac Volts
Test instrument
Calibrator output indications (V ac)
Range
Indication
Frequency
Min
Max
pushbutton
2
1.999
50 Hz
1.9790
2.0190
2
1.999
1 kHz
1.9790
2.0190
2
1.999
10 kHz
1.9790
2.0190
2
1.999
20 kHz
1.9510
2.0470
2
1.999
100 kHz
1.8870
2.1110
15