Frequency Response
Distortion

Calibration Procedure For Test Oscillator, Models 651A, 651B And 652A (Sg-763U) Page Navigation
  5    6    7    8    9  10  11    12    13    14  
img
. . . . .
TB 9-6625-1356-24
setting. If multimeter does not indicate within limits recorded in (9) through (11) above for
+ (plus) and - minus perform b below.
CAUTION
Adjust AMPLITUDE controls for minimum output when
changing FREQUENCY RANGE switch setting.
Table 5. Frequency Response
Test instrument
Multimeter indications
FREQUENCY RANGE
FREQUENCY
Hewlett-Packard
All other
switch settings
dial settings
model 651A
models
3%
2%
X10
2
3%
2%
X10
1
2%
1%
X100
1
2%
1%
X1K
5
2%
1%
X10K
5
2%
1%
X100K
5
2%
1%
X1M
1
4%
2%
X1M
4
4%
2%
X1M
6
4%
2%
X1M
8
4%
2%
X1M
10
b. Adjustments
(1) Adjust A2C14 (fig. 1) fm best overall flatness between 1 and 10 MHz.
(2) Set FREQUENCY RANGE switch to X1M and turn FREQUENCY dial to 10.
(3) Adjust AMPLITUDE controls for multimeter indication recorded in a (5) above.
(4) Adjust A1C15 (fig. 1) for a 2.40 V indication on TI meter (R).
11. Expand Monitor Flatness (Model 652A (SG-763/U) Only)
a. Performance Check
(1) Connect multimeter to TI using a 50  feedthrough termination.
(2) Position controls as listed in (a) through (f) below:
FREQUENCY RANGE switch to X100.
(a)
FREQUENCY dial to 4.
(b)
AMPLITUDE controls for a 3.00 V indication on multimeter.
(c)
NORMAL-EXPAND switch to EXPAND.
(d)
REF SET control for a 0 indication on TI meter expand scale.
(e)
NOTE
Do not change REF SET controls throughout remainder of test.
(f) AMPLITUDE controls for minimum output.
(3) Substitute thermal converter for 50  feedthrough termination.
Adjust
AMPLITUDE controls for 0 indication on TI meter expand scale.  Record multimeter
indication.
11