TB 9-6625-1356-24
Test instrument
FREQUENCY RANGE switch settings
FREQUENCY dial settings
X10
5
X10
10
X100
1
X100
10
X1K
10
X10K
1
X10K
2
X100K
1
b. Adjustments. Adjust A2R17 (fig. 1) for minimum distortion indication on audio
analyzer (R).
13. Output Attenuator Accuracy
a. Performance Check
(1) Adjust AMPLITUDE controls for minimum output.
RMS 3.173 V AC
Figure 3. Output attenuator - equipment setup.
(3) Position controls as listed in (a) through (c) below:
(c) OUTPUT ATTENUATOR switch to +20 DBM.
(4) Adjust attenuator to 60 dB.
(5) Adjust AMPLITUDE controls for a 1 mV reference indication on true rms
(6) Set OUTPUT ATTENUATOR switch to +10 DBM.
Change 1