TB 9-6625-143-24
IDENTIFICATION AND DESCRIPTION
This bulletin provides instructions for the
1. Test Instrument Identification.
calibration of Oscilloscope, Agilent, Models 54622A, 54622D and 54624A.
The
manufacturer's manual was used as the prime data sources in compiling these instructions.
The equipment being calibrated will be referred to as the TI (test instrument) throughout
this bulletin.
a. Model Variations. Variations among models are described in text and tables.
b. Time and Technique. The time required for this calibration is approximately 2
hours, using the dc and low frequency technique.
2. Forms, Records, and Reports
a. Forms, records, and reports required for calibration personnel at all levels are
prescribed by TB 750-25.
b. Adjustments to be reported are designated (R) at the end of the sentence in which
they appear. When adjustments are in tables, the (R) follows the designated adjustment.
Report only those adjustments made and designated with (R).
3. Calibration Description. TI parameters and performance specifications which
pertain to this calibration are listed in table 1.
Test instrument parameters
Performance specifications
Threshold accuracy
Range:
-8 V to +8 V
(Model 54622D only)
Accuracy: (100 mV + 3% threshold setting)
DC vertical gain
Range:
1 mV to 5 V/div
Accuracy: 2% FS1
Vertical dual cursor
Range:
1 mV to 5 V/div
Accuracy: (DC vertical gain accuracy + 0.4% FS (~1 LSB))1, 2
Range:
DC to 100 MHz, DC Coupled
Accuracy: Down not more than 3 dB
Edge triggering
Range:
Internal: 6 div;
Sensitivity: greater of 0.35 div or 2.5 mV
External: <75 mV p-p @ 25 MHz; <150 mV p-p @ 100 MHz
Accuracy: Stable display
Delta time
Range:
5 ns to 50 s/div
Accuracy: Same Channel:
(0.01% rdg + 0.1% screen width + 40 ps)
1 mV/div is a magnification of 2 mV/div. For TI vertical accuracies, use FS of 16 mV for 1 mV/div setting.
1
1 LSB = 0.4% FS.
2