SECTION IV. CALIBRATION PROCESS FOR VOLTAGE DIVIDER FLUKE MODEL 80E-10
Figure 7. Divider accuracy.

Calibration Procedure For Differential Voltmeter Models 887A, 887Ab, And 887Ab/An, And Voltage Divider Model 80E-10 Page Navigation
  14    15    16    17    18  19  20    21  
img
. . . . .
TB 9-6625-1493-24
17. Equipment Setup
WARNING
HIGH VOLTAGE is used or exposed during the performance of
this calibration.
DEATH ON CONTACT may result if
personnel fail to observe safety precautions.
REDUCE
OUTPUT(S) TO MINIMUM after each step within the
performance check where applicable.
a. Remove protective cover from TI only when necessary to make adjustments. Replace
cover after completing the adjustments.
b. Adjust TI mechanical zero adjustment for 0 indication on KILOVOLTS meter.
c. Connect TI to 0-10-KV OUTPUT jack on dc power supply.
d. Adjust dc power supply for a 10,000 V output and allow 30 minutes for TI to stabilize.
18. Meter Accuracy
a. Performance Check
(1) Adjust dc power supply for an indication of 8 on TI KILOVOLTS meter. If dc
power supply does not indicate between 7700 and 8300 V, perform b below.
(2) Reduce dc power supply output to minimum.
b. Adjustments
(1) Adjust dc power supply for a 10,000 V output.
(2) Adjust R4 (fig. 6) for TI indication of 10 (R).
Figure 6. Voltage divider - adjustment locations.
19