Differential Voltmeter AC Voltage Accuracy (Reference Level)
Table 8. Ac Voltage (Reference Level)

Calibration Procedure For Differential Voltmeter Models 887A, 887Ab, And 887Ab/An, And Voltage Divider Model 80E-10 Page Navigation
  10    11    12    13    14  15  16    17    18    19    20  
img
. . .
TB 9-6625-1493-24
(5) Measure ac voltage. If AC measurement standard indication is not between
.999445 and 1.000555 V dc, perform b below.
(6) Repeat technique of (3) through (5) above for calibrator initial outputs and TI
setting listed in table 8. If AC measurement standards are not within limits specified,
perform corresponding adjustments, using technique of b below.
CAUTION
Table 8 requires progressive increase in voltage output from
calibrator. To avoid damage to TI, ensure TI is set to proper
range before increasing the output from calibrator.
Figure 4. Ac voltage (reference level).