RF Spectrum Analyzer - Continued
Figure 8. Power Meter Connection.

Calibration Procedure For Radio Test Set, Models 2947 And 2947A And Model 2947 Page Navigation
  35    36    37    38    39  40  41    42    43    44    45  
img
. .
TB 9-6625-152-24
(107) Press TI [Return], [Scope/Bar] and [Scope] keys. Set TI timebase to 500 μs,
] and [
] keys.
using [
(108) Press TI [Return] and [Mod Meter] keys. Use [AM/FM/SSB] key to select
SSB.
(109) Use TI [RS Sens] key to select High Sens.
(110) Check that TI receiver section is correctly tuned. This is shown by TI frequency
offset indicating 0 Hz 10 Hz, and indicated power of -90 dBm 3 dB.
(111) Press TI [Return], [Tx Freq] and 580.001 MHz keys.
(112) Turn TI VOLUME control clockwise and ensure a 1 kHz tone is audible.
(113) Observe TI oscilloscope display and adjust R24 (fig. 7) for 3 vertical divisions of
waveform.
Figure 7. A15 SSB Board.
40