SECTION IV. ADJUSTMENT PROCESS
Figure 10. TI Board Placement.

Calibration Procedure For Radio Test Set, Models 2947 And 2947A And Model 2947 Page Navigation
  43    44    45    46    47  48  49    50    51    52    53  
img
. .
TB 9-6625-152-24
NOTE
Each adjustment procedure which requires access to the
diagnostic or calibrate menus shows the full keypress sequence
commencing with the [HELP/SETUP] key. If adjustment
instructions are followed consecutively, a diagnostic or
calibrate menu will continue to be displayed and there will be
no need to use the full key press sequence.  When making
consecutive adjustments, ignore key sequence inside braces { }.
NOTE
If 600
option is installed, set TI Audio Input Impedance to
high, Audio Output Impedance to low, and 20 dB Audio
Attenuator to 0 dB.
NOTE
Adjustments in the following paragraphs are interactive. Upon
completion of adjustments in this section, performance of
paragraphs 15a, 16a, 17a, 18a, 19a, 20a and 21a will need to
be completed.
NOTE
For board placement see figure 10 below.
48