Table 21. Edge 150 ps Edge Gain.
Table 23. Model 9560 Edge Rise Time Accuracy.

Calibration Procedure For Calibrator Active Head, Models 9530, 9550 And 9560 Page Navigation
  13    14    15    16    17  18  19    20    21    22    23  
img
. . .
TB 9-6625-154-40
(29) Set TI as listed in (a) through (i) below:
(a)
OUTPUT OFF.
(b)
Softkey, EXIT.
(c)
Softkey, STORE HEAD CAL.
(d)
Correct all necessary Cal Date data.
(e)
Softkey, STORE.
(f)
Wait for Saving cal factors to head to finish then press Softkey, EXIT.
(g)
Softkey, EXIT.
(h)
MODE.
(i)
Softkey, MANUAL.
(30) Disconnect setup and repeat paragraph 10 a above.
11. Edge Rise Time Accuracy (Model 9560 Only)
a. Performance Check
(1) Connect TI output to storage oscilloscope input.
(2) Connect TI CH5 output to storage oscilloscope trigger input.
(3) Set storage oscilloscope to measure rise time with full bandwidth.
(4) Set TI output as listed in (a) through (o) below:
(a)
Softkey,  / 1.0 (select 1.0)
(b)
Softkey, CHANNEL SELECT.
Softkey, LOAD 50 Ω.
(c)
(d)
Softkey, SIGNAL CH 1.
(e)
Softkey, EXIT.
(f)
(Edge).
(g)
Softkey,
FAST.
(h)
Softkey,
70 ps.
(i)
(if necessary) to select AMPLITUDE.
(j)
Keypad Entry, 33.
(k)
Softkey, mV.
(l)
(if necessary) to select FREQUENCY.
(m)
Keypad Entry, 1.
(n)
Softkey, MHz.
(o)
OUTPUT, ON.
(5) Set TI CH 5 (Trigger channel) output as listed in (a) through (g) below:
(a)
Softkey, CHANNEL SELECT.
(b)
Softkey, TRIGGER CHANNEL.
(c)
Softkey, TRIGGER CHANNEL 5.
(d)
Softkey, CABLE SELECT.
(e)
Softkey, CABLE CH 5.
(f)
Softkey, EXIT.
(g)
Softkey, EXIT.
(6) Adjust the storage oscilloscope for a full scale rise time display with full
bandwidth.
19