Table 18. Edge 500 ps Edge Gain.
Table 21. Edge 150 ps Edge Gain.

Calibration Procedure For Calibrator Active Head, Models 9530, 9550 And 9560 Page Navigation
  11    12    13    14    15  16  17    18    19    20    21  
img
. . . . .
TB 9-6625-154-40
Table 19. Edge 500 ps Edge Speed.
Measured
Calculated
Frequency
TI level
Cal Point
Edge
Edge Time
Adjustment Value
(MHz)
(V p-p)
(ps)
(ps)
Target 1
1
3.0000
Rising
1
Target 2
2.0000
Rising
1
Target 3
1.0000
Rising
1
Target 4
1.0072
Rising
1
Target 5
316.49 m
Rising
1
Target 6
316.49 m
Rising
1
Target 7
3.0000
Falling
1
Target 8
2.0000
Falling
1
Target 9
1.0000
Falling
1
Target 10
1.0072
Falling
Target 11
1
316.49 m
Falling
1
Target 12
316.49 m
Falling
(16) Set TI as listed in (a) through (f) below:
(a)
Softkey, EXIT.
(b)
Softkey, Edge.
(c)
Softkey, 150 ps Edge.
(d)
Softkey, LIN.
(e)
Softkey, TARGET 1.
(f)
OUTPUT, ON.
(17) Calculate the adjustment value using the following formula and record in
table 20.
Adjustment value = TI level(Vp-p) *
TRV = Test reported value for the attenuator placed between
. = 0.1023293
TI and storage oscilloscope.
Example: Adjustment value = 1.0000 *
(18) Using the TI arrow keys adjust the TI O/P Amplitude until the storage
oscilloscope amplitude display matches the first adjustment value listed in table 20 below
and press TI Softkey, ACCEPT CALIB.
(19) Repeat technique of (16)(e) through (18) above for remaining rows in table 20
adjusting the appropriate edge.
Table 20. Edge 150 ps Edge Linearity.
Frequency
TI level
Adjustment Value
Edge
Cal Point
(kHz)
(V p-p)
(V p-p)
Target 1
100
1.0000
Rise
2.0000
Target 2
100
Rise
100
2.5000
Target 3
Rise
100
Target 4
1.0000
Fall
100
2.0000
Target 5
Fall
2.5000
Target 6
100
Fall
17