TB 9-6625-154-40
Table 12. 3 GHz Leveled Sine Function HF Linearity.
Model 9560
Model 9530
TI level
Power Meter
TI level
Power Meter
(V p-p)
Indication
(V p-p)
Indication
Cal Point
-2.956
Target 1
450.00 m
-2.956
450.00 m
800.00 m
700.00 m
0.881
Target 2
2.041
1.2000
1.0000
3.979
Target 3
5.563
2.0000
1.6000
8.062
Target 4
10.000
Target 5
2.8000
2.4000
11.584
12.923
Target 6
3.7000
3.5000
14.861
15.343
(17) Set TI as listed in (a) through (c) below:
(a) Softkey, FLAT.
(b) Softkey, TARGET 1.
(c) OUTPUT ON.
(19) Using the TI arrow keys adjust the TI O/P Amplitude until the power meter
display matches the first Meter Indication listed in table 13 below and press TI Softkey,
ACCEPT CALIB.
(20) Repeat technique of (17)(b) and (19) above for remaining rows in table 13.
(21) Press TI Softkey, NEXT FREQ and repeat technique of (17)(b) through (20)
above for 1.5 GHz.
(22) Repeat technique of (21) above for 1.75, 1.95, 2.05, 2.2, 2.3, 2.45, 2.55, 2.65, 2.80,
2.90, 3.00, 3.10, and 3.20 GHz.
Table 13. 3 GHz Leveled Sine Function HF Flatness.
Model 9560
Model 9530
Cal Point
TI level (V p-p)
Power Meter (dBm)
TI level (V p-p)
Power Meter (dBm)
1.6000
Target 1
2.0000
10.000
8.062
671.48 m
526.30 m
Target 2
0.520
-1.596
210.99 m
155.30 m
Target 3
-9.535
-12.197
70.839 m
51.300 m
Target 4
-19.015
-21.818
Target 5
22.259 m
15.240 m
-29.071
-32.361
NOTE
Techniques of (23) through (32) below
apply to model 9560 only.
(23) Set TI as listed in (a) through (e) below:
(a)
Softkey, EXIT.
(b)
Softkey, SINE.
(c)
Softkey, 6.4 GHz Sine.
(d)
Softkey, TARGET 1.
(e)
OUTPUT ON.
(25) Using the TI arrow keys adjust the TI O/P Amplitude until the power meter
display matches the first Meter Indication listed in table 14 below and press TI Softkey,
ACCEPT CALIB.
(26) Repeat technique of (23)(d) and (25) above for remaining rows in table 14.
12
Change 2