Table 6. 3 V Leveled Sine Flatness Accuracy.
Table 9. 100 mV Leveled Sine Flatness Accuracy.

Calibration Procedure For Calibrator Active Head, Models 9530, 9550 And 9560 Page Navigation
  3    4    5    6    7  8  9    10    11    12    13  
img
. . . .
TB 9-6625-154-40
Table 7. 1 V Leveled Sine Flatness Accuracy.
Freq (Hz)
Min (dB)
Measured
Max (dB)
50 k
REF
REF
10 M
-0.175
+0.172
50 M
-0.175
+0.172
100 M
-0.175
+0.172
250 M
-0.175
+0.172
400 M
-0.220
+0.214
550 M
-0.220
+0.214
600 M
-0.220
+0.299
725 M
-0.309
+0.299
1G
-0.309
+0.299
1.5 G
-0.355
+0.341
2G
-0.355
+0.341
2.5 G
-0.355
+0.341
3.0 G
-0.355
+0.341
1
4.0 G
-0.355
+0.341
5.0 G
-0.355
+0.341
5.5 G
-0.355
+0.341
6.0 G
-0.355
+0.341
Remaining rows apply to model 9560 only.
1
(13) Repeat technique of (4) through (7) above for values in table 8. TI will indicate
within the limits of table 8.
Table 8. 300 mV Leveled Sine Flatness Accuracy.
Freq (Hz)
Min (dB)
Measured
Max (dB)
50 k
REF
REF
10 M
-0.175
+0.172
50 M
-0.175
+0.172
100 M
-0.175
+0.172
250 M
-0.175
+0.172
400 M
-0.220
+0.214
550 M
-0.220
+0.214
600 M
-0.220
+0.299
725 M
-0.309
+0.299
1G
-0.309
+0.299
1.5 G
-0.355
+0.341
2G
-0.355
+0.341
2.5 G
-0.355
+0.341
3.0 G
-0.355
+0.341
1
4.0 G
-0.355
+0.341
5.0 G
-0.355
+0.341
5.5 G
-0.355
+0.341
6.0 G
-0.355
+0.341
Remaining rows apply to model 9560 only.
1
(14) De-select relative measure mode on power meter.
(15) Set TI as listed in (a) through (f) below:
(a)
(if necessary) to select AMPLITUDE.
(b)
Keypad Entry, 100.
(c)
Softkey, mV.
(d)
(if necessary) to select FREQUENCY.
(e)
Keypad Entry, 50.
(f)
Softkey, kHz.