Table 1. Calibration Description Continued.
Equipment Setup

Calibration Procedure For Calibrator Active Head, Models 9530, 9550 And 9560 Page Navigation
  1    2  3  4    5    6    7    8  
img
. . . . . .
TB 9-6625-154-40
Table 2. Minimum Specifications of Equipment Required.
Manufacturer and model
Common name
Minimum use specifications
(part number)
ATTENUATOR FIXED
20 dB verified
Agilent, Model 8493C-020
or Weinschel, Model 9918-
20dB (9918)
MULTIMETER
ACV
Fluke, Model 8508A01
Range: 10 mV to 1 V
(8508A/01) or Hewlett
Frequency: 1k Hz to 50 kHz
Packard, Model 3458A
Accuracy:
(3458A)
1 to 45 kHz
0.5 %
50 kHz
0.375 %
Rise Time ≤ 17.5 ps
OSCILLOSCOPE
Agilent, Model 86100C w/
Frequency range: 100 kHz to 1 MHz
plug-in module 86112A
Accuracy: capability
Amplitude range: 106 to 750 mV
Accuracy: (0.4%FS + 2mV + 1.5%
CH Offset)
OSCILLOSCOPE
Calibrated
Fluke 9500B/3200
CALIBRATOR
POWER METER
Frequency: 10 MHz to 6.4 GHz
Agilent, Model E4419B p/o
Power Level: -10.0 to 18.7867 dBm
Measuring receiver system
Accuracy:
N5530S (E4419B) with
10 MHz
1.0 %
w/power sensor
50 to 250 MHz
1.0 %
E9304AH18H84
400 MHz
1.25 %
550 MHz
1.25 %
600 MHz to 1 GHz
1.75%
Or
1.5 to 6.4 GHz
2.0%
Power Level: -16.0206 to -10 dBm
Agilent, Model N1911A
Accuracy:
(N1911A) with w/power
10 MHz
1.0 % (1.5 %)
sensor E9304AH18H84
50 to 250 MHz
1.0 % (1.5 %)
400 MHz
1.25 % (1.5 %)
550 MHz
1.25 % (1.5 %)
600 MHz to 1 GHz
1.75% (1.5 %)
1.5 to 6.4 GHz
2.0%
SECTION III
CALIBRATION PROCESS
6. Preliminary Instructions
a. The instructions outlined in paragraphs 6 and 7 are preparatory to the calibration
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
name as listed in table 2.
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure.  Additional
maintenance information is contained in the manufactures manual for this TI.
d. Unless otherwise specified, all controls and control settings refer to the TI.