TB 9-6625-154-40
IDENTIFICATION AND DESCRIPTION
This bulletin provides instructions for the
1. Test Instrument Identification.
calibration of Active Head, Fluke, Models 9530, 9550 and 9560. The manufacturer's
manual was used as the prime data source in compiling these instructions. The equipment
being calibrated will be referred to as the TI (test instrument) throughout this bulletin.
a. Model Variations. Variations among models are described in text and tables.
b. Time and Technique. The time required for this calibration is approximately 8
hours, using the dc and low frequency technique.
2. Forms, Records, and Reports. Forms, records, and reports required for calibration
personnel at all levels are prescribed by TB 750-25.
3. Calibration Description. TI parameters and performance specifications which
pertain to this calibration are listed in table 1.
Test instrument parameters
Performance specifications
EDGE PULSE RISE/FALL TIME
Model 9530
Range: 500 ps
Accuracy: +50 to -150 ps
Accuracy of displayed value 40 ps
Range: 150 ps
Accuracy: 25 ps
Accuracy of displayed value 15 ps
Model 9550
Range: 25 ps
Accuracy: 4 ps
Accuracy of displayed value 4 ps
Model 9560 (rise time only)
Range: 70 ps
Accuracy: +15ps to -20 ps
Accuracy of displayed value 11 ps
SINE AMPLITUDE AND FLATNESS
Model 9530
Range: 0.1 Hz to 3.2 GHz
Amplitude (p-p)
Range: 0.1 Hz to 550 MHz
5 mV to 5 V
Range: 550 MHz to 2.5 GHz
5 mV to 3 V
Range: 2.5 GHz to 3.2 GHz
5 mV to 2 V
Flatness Accuracy:
Range: 0.1 Hz to 300 MHz
Accuracy: 2%
Range: 300 MHz to 550 MHz
Accuracy: 2.5%
Range: 550 MHz to 1.1 GHz
Accuracy: 3.5%
Range: 1.1 GHz to 3.2 GHz
Accuracy: 4%
Absolute Accuracy
Range: 50 kHz to 10 MHz
Accuracy: 1.5%