Edge Rise Time Accuracy (Model 9560 Only)
Table 24. Edge 70 ps Edge Linearity.

Calibration Procedure For Calibrator Active Head, Models 9530, 9550 And 9560 Page Navigation
  14    15    16    17    18  19  20    21    22    23    24  
img
. . . . . . .
TB 9-6625-154-40
(7) Storage oscilloscope mean display will indicate within the storage oscilloscope
limits of table 23. If not, perform b below.
(8) The difference between the measured value and the TI displayed value will be
within TI Display limits listed in table 23. If not, perform b below.
(9) Repeat technique of (4)(j) through (4)(n) and (6) through (8) above for remaining
TI frequencies in table 10. TI will indicate within the limits of table 23.
Table 23. Model 9560 Edge Rise Time Accuracy.
TI
Storage oscilloscope
TI Display
Rise/Fall
Output
Lower Limit
Upper Limit
Limit
Edge Type
(ps)
(V p-p)
(ps)
(ps)
(ps)
70
Pos
33 m
11
46.9511389
96.57536453
70
Pos
66 m
11
46.9511389
96.57536453
70
Pos
100 m
11
46.9511389
96.57536453
70
Pos
200 m
11
46.9511389
96.57536453
70
Pos
330 m
11
46.9511389
96.57536453
70
Pos
660 m
11
46.9511389
96.57536453
70 1
Pos
1
11
46.9511389
96.57536453
70
Pos
2
11
46.9511389
96.57536453
Insert a 20 dB attenuator between the TI and the storage oscilloscope input.
1
(10) Set TI output to OFF and disconnect storage oscilloscope.
b. Adjustments
NOTE
Ensure the 9500B rear-panel CAL ENABLE/DISABLE switch
is in the ENABLE position and return to original position
when complete.
(1) Connect TI output to the storage oscilloscope input.
(2) Connect TI CH5 output to storage oscilloscope trigger input.
(3) Set storage oscilloscope to measure edge and amplitude V p-p with full
bandwidth.
(4) Set TI as listed in (a) through (h) below:
(a)
MODE.
(b)
Softkey, CALIB.
Enter Password (exp. 2→3→5→7 ).
(c)
(d)
Softkey, HEAD CAL.
(e)
Softkey, Edge.
(f)
Softkey, 70 ps Edge.
(g)
Softkey, LIN.
(h)
OUTPUT, ON.
(5) Using the TI arrow keys adjust the TI O/P Amplitude until the storage
oscilloscope amplitude display matches the first adjustment value listed in table 24 below
and press TI Softkey, ACCEPT CALIB.
(6) Press Softkey, NEXT TARGET and repeat technique of (5) above for remaining
rows in table 24 adjusting the appropriate edge.