TB 9-6625-157-40
IDENTIFICATION AND DESCRIPTION
1. Test Instrument Identification.
This bulletin provides instructions for the
calibration of Piston Gage Base, Fluke, models PG7202, PG7302 and PG7601. The
manufacturer's manual was used as the prime data source in compiling these instructions.
The equipment being calibrated will be referred to as the TI (test instrument) throughout
this bulletin.
a. Model Variations. Differences in models are described in text and tables.
b. Time and Technique. The time required for this calibration is approximately 3
hours, using the physical and dimensional technique.
2. Forms, Records, and Reports
a. Forms, records, and reports required for calibration personnel at all levels are
prescribed by TB 750-25.
b. Adjustments to be reported are designated (R) at the end of the sentence in which
they appear. When adjustments are in tables, the (R) follows the designated adjustment.
Report only those adjustments made and designated with (R).
3. Calibration Description. TI parameters and performance specifications which
pertain to this calibration are listed in table 1.
Performance specifications1
Test instrument parameters
Ambient Temperature
Range:
0 to 40 C
Accuracy: 1 C
Relative Humidity
Range:
5 to 95 % RH
Accuracy: 10 % RH
Atmospheric Pressure
Range:
70 to 110 kPa
Accuracy: 140 Pa
Piston Cylinder Temperature
Range:
0 to 40 C
Accuracy: 0.1 C or 10% of reading, whichever is greater
Vacuum (PG7601 Only)
Range:
0 to 20 Pa
Accuracy: 0.1 Pa or 10% of reading, whichever is greater
1 Single point calibration under ambient conditions only.