TB 9-6625-157-40
(b) <7cal>
(c) <3%RH>
(d) <1view>
(4) The TI will display the output in the format shown in figure 2.
Current measurement of
Indication that this is a
the Relative Humidity
Relative Humidity
Sensor
Display
43.2
%RH
%RH
HA 5 %RH
HM
0.99984
Current value of the
Current value of the
Humidity Adder
Humidity Multiplier
(HA)
(HM)
(5) Compare TI %RH indication to monitor h-t indication. TI and monitor h-t will
agree within 10% RH. If not, perform b below.
b. Adjustments
(1) Calculate and record the difference between TI %RH indication and monitor h-t
using the formula below:
Monitor h-t [%RH] %RH [%RH] = Ć%RH
(2) Record the current value of the HA (Humidity Adder) from the TI display
(3) Calculate the New HA by algebraically adding the Ć%RH to the Current HA:
Current HA [%RH] + Ć%RH [%RH] = New HA [%RH]
(4) Adjust HA by pressing TI keys as listed in (a) through (f) below. (R)
[ESCAPE]
(a)
<2cal>
(b)
Using key pad, input new HA.
(c)
[ENTER]
(d)
[ENTER]
(e)
<2yes>
(f)
6