Piston Cylinder Temperature
Final Procedure

Calibration Procedure For Piston Gage Base Sensors, Models Pg7202, Pg7302 And Pg7601 Page Navigation
  5    6    7    8    9  10  11  
img
. . .
TB 9-6625-157-40
(6) Read PCT (current measurement of the piston cylinder temperature sensor) by
pressing TI keys as listed in (a) through (d) below.
[SPECIAL]
(a)
<7cal>
(b)
<4PCT>
(c)
<1view>
(d)
(7) The TI will display the output in the format shown in figure 6.
Current measurement of
Indication that this is a
the Piston Cylinder
Piston Cylinder
Temperature Sensor
Temperature Display
19.951 C
PCT
RZ 100.003 Ω
S 0.3896
Current Resistance
Current value of the
Offset at 0C in
Resistance Slope
Ohms (RZ)
(S)
Figure 6. Piston Cylinder Temperature Display.
(8) Wait approximately 60 minutes for temperature readout to stabilize.
NOTE
If total temperature change over the time of the test of either
the Thermometer or PCT is greater than 0.025 C, the
temperature instability during the test is excessive and the
measurements should be repeated.
(9) Record the current reading of PCT and the thermometer readout every
30 seconds for 5 minutes.
(10) Compare TI PCT indications to thermometer readout indications. TI and
thermometer readout will agree within 0.1 C. If not, perform b below.
b. Adjustments
(1) Average the TI PCT indications and thermometer readout indications from
paragraph 12 a (8).
(2) Calculate and record the difference between TI PCT average and temperature
readout average using the formula below:
Thermometer readout Avg. [C] PCT Avg. [C] = ĆAvg. Temp [C]