Figure 6. Piston Cylinder Temperature Display.

Calibration Procedure For Piston Gage Base Sensors, Models Pg7202, Pg7302 And Pg7601 Page Navigation
  6    7    8    9    10  11
img
. . . .
TB 9-6625-157-40
(3) Calculate the resistance corresponding to ĆAvg. Temp using formula below:
ĆAvg. Temp [C] x 0.3896 [Ω/C] = ĆResistance [Ω]
(4) Record the current value of the RZ (current resistance offset) from the TI
display (fig. 6).
(5) Calculate the new RZ by algebraically adding the ĆResistance to the current RZ
using formula below:
Current RZ [Ω] + ĆResistance [Ω] = New RZ [Ω]
(6) Adjust RZ by pressing TI keys as listed in (a) through (f) below. (R)
[ESCAPE]
(a)
<2cal>
(b)
Using key pad, input new RZ.
(c)
[ENTER]
(d)
[ENTER]
(e)
<2yes>
(f)
(7) Repeat paragraph 12 a.
(8) Remove thermometer and PRT insertion plug from TI.
(9) Replace the piston cylinder module, mass set and automatic mass handler.
13. Final Procedure
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.