Figure 3: Atmospheric Pressure Sensor Display.
Figure 5. Vacuum Sensor Display.

Calibration Procedure For Piston Gage Base Sensors, Models Pg7202, Pg7302 And Pg7601 Page Navigation
  2    3    4    5    6  7  8    9    10    11  
img
. . . . .
TB 9-6625-157-40
(5) Adjust PA by pressing TI keys as listed in (a) through (f) below. (R)
[ESCAPE]
(a)
<2cal>
(b)
Using key pad, input new PA.
(c)
[ENTER]
(d)
[ENTER]
(e)
<2yes>
(f)
(6) Repeat paragraph 10 a.
11. Vacuum (Model PG7601 Only)
a. Performance Check
(1) Place mass set in the load all configuration as listed in (a) through (e) below:
(a) [SPECIAL]
(b) <8AMH>
(c) <2control>
(d) <3loadall>
(e) [ENTER]
(2) Remove the automatic mass handler, mass set, and piston cylinder module from
the TI.
(3) Place automatic mass handler and bell jar back on the piston gage base.
(4) Install the vacuum standard with readout and accessories onto the TI as shown
in figure 4.
Figure 4. Vacuum Setup.
(5) Connect readout to vacuum standard.
(6) Connect power cord to readout.
(7) Convert readout from Torr to Pa (1 Torr = 133.32 Pa) by pressing the readout
keys as listed in (a) through (e) below.