SECTION III. CALIBRATION PROCESS
Table 3. Model 2608A Ac Current.

Calibration Procedure For Clamp Meters Model 2608A And Models 380911F, 380912, 380920, 380947 And 380973 Page Navigation
  1    2    3    4    5  6  7    8    9    10    11  
img
. . . . . . . . . . . . . . . . . .
TB 9-6625-159-24
7. Equipment Setup
WARNING
HIGH VOLTAGE is used or exposed during the performance of
this calibration.
DEATH ON CONTACT may result if
personnel fail to observe safety precautions.
REDUCE
OUTPUT(S)  to  minimum  after  each  step  within  the
performance check where applicable.
(1) If applicable, slide TI power switch to ON and ensure DATA HOLD button is OFF.
(2) Zero TI range/function as necessary.
8. Ac Current
a. Performance Check
(1) Connect calibrator to current coil.
(2) Set TI function switch to A ac.
(3) Position TI clamp so it is centered on current coil.
(4) Set TI and calibrator to settings listed in table 3 through table 8 for appropriate
TI model. TI will indicate within limits specified.
b. Adjustments. Refer to TI manufacturer's manual for corrective action.
9. Ac Voltage
a. Performance Check
(1) Set TI function switch to V ac.
(2) Set TI and calibrator to settings listed in table 9 through table 13 for
appropriate TI model. TI will indicate within limits specified.
b. Adjustments. Refer to TI manufacturer's manual for corrective action.
10. Dc Voltage
a. Performance Check
(1) Set TI function switch to V dc.
(2) Set TI and calibrator to settings listed in table 14 through table 18 for
appropriate TI model. TI will indicate within limits specified.
b. Adjustments. Refer to TI manufacturer's manual for corrective action.
7