Table 12. Model 380920 Ac Voltage.
Resistance

Calibration Procedure For Clamp Meters Model 2608A And Models 380911F, 380912, 380920, 380947 And 380973 Page Navigation
  6    7    8    9    10  11  12    13    14    15    16  
img
.
TB 9-6625-159-24
Table 14. Model 2608A Dc Voltage.
Performance Checks
Test instrument indications
Test instrument
Calibrator
range settings
output settings
Min
Max
(V dc)
(V dc)
(V dc)
(V dc)
60
20
18.2
21.8
60
30
28.2
31.8
60
40
38.2
41.8
60
50
48.2
51.8
60
60
58.2
61.8
Table 15. Model 380911F Dc Voltage.
Performance Checks
Test instrument
Calibrator
Test instrument indications
range settings
output settings
Min
Max
(V dc)
(V dc)
(V dc)
(V dc)
1000
300
294
306
1000
700
688
712
1000
950
935
965
1000
-950
-965
-935
Table 16. Model 380912 Dc Voltage.
Performance Checks
Test instrument
Calibrator
Test instrument indications
range settings
output settings
Min
Max
(V dc)
(V dc)
(V dc)
(V dc)
200
60
59.4
60.6
200 1
100
97.0
103.0
200
140
138.8
141.2
200
180
178.5
181.5
200
-180
-181.8
-178.5
1000
900
892
908
Set TI Peak Hold on and calibrator to STBY. Check limits then set TI Peak Hold off.
1
Table 17. Model 380920 Dc Voltage.
Performance Checks
Test instrument
Calibrator
Test instrument indications
range settings
output settings
Min
Max
(V dc)
(V dc)
(V dc)
(V dc)
400 m
360  m
354.3  m
365.7  m
4
3.6
3.543
3.657
40
12
11.79
12.21
40
28
27.55
28.45
40
36
35.43
36.57
40
-36
-36.57
-35.43
400
360
354.3
365.7
600
540
529.0
551.0
12