TB 9-6625-159-24
Table 11. Model 380912 Ac Voltage.
Performance Checks
Calibrator output
Test instrument indications
Test instrument
settings
range settings
Voltage
Min
Max
(V ac)
(V ac)
(Hz)
(V ac)
(V ac)
200
180
60
178.0
182.0
750
675
60
666
684
Table 12. Model 380920 Ac Voltage.
Performance Checks
Calibrator output
Test instrument indications
Test instrument
settings
range settings
Voltage
Min
Max
(V ac)
(V ac)
(Hz)
(V ac)
(V ac)
4
3.6
45
3.523
3.677
4
3.6
60
3.541
3.659
4
3.6
1k
3.523
3.677
40
12
45
11.71
12.29
40
28
60
27.53
28.47
40
36
1k
35.23
36.77
400
360
45
352.3
367.7
400
360
60
354.1
365.9
400
360
1k
352.3
367.7
600
540
45
524
556
600
540
60
527
553
600
540
1k
524
556
Table 13. Model 380973 Ac Voltage.
Performance Checks
Calibrator output
Test instrument indications
Test instrument
settings
range settings
Voltage
Min
Max
(V ac)
(V)
(Hz)
(Hz)
(Hz)
400
360
60
355.2
364.8
400 1
360
60
353.6
366.4
400
360
400
355.2
364.8
600
540
60
529
551
600
360
400
529
551
Set TI Peak Hold on and calibrator to STBY. Check limits then set TI Peak Hold off.
1