• Home
  • Download PDF
  • Order CD-ROM
  • Order in Print
Resistance
Table 23. Model 380973 Resistance.

Calibration Procedure For Clamp Meters Model 2608A And Models 380911F, 380912, 380920, 380947 And 380973 Page Navigation
  8    9    10    11    12  13  14    15    16    17  
img
. .
TB 9-6625-159-24
Table 19. Model 2608A Resistance.
Performance Checks
Resistance standard
Test instrument
Output
Range
Indication limits
(Ω)
(Ω)
Min (Ω)
Max (Ω)
10
X1
9.8
10.2
20
X1
19.6
20.4
50
X1
49.0
51.0
100
X1
98.0
102.0
1 k
X1
980.0
1020.0
1
100
X10
98.0
102.0
1
200
X10
196.0
204.0
X101
500
490
510
1
1 k
X10
980.0
1020.0
1
10 k
X10
9800.0
10200.0
1
Multiply reading by 10.
Table 20. Model 380911F Resistance.
Performance Checks
Resistance standard
Test instrument
Output
Range
Indication limits
(Ω)
(Ω)
Min (Ω)
Max (Ω)
1.9 k
2k
1.888 k
1.932 k
190  k
200 k
186.8  k
193.2  k
Table 21. Model 380912 Resistance.
Performance Checks
Resistance standard
Test instrument
Output
Range
Indication limits
(Ω)
(Ω)
Min (Ω)
Max (Ω)
1800
2000
1781
1819
Table 22. Model 380920 Resistance.
Performance Checks
Resistance standard
Test instrument
Output
Range
Indication limits
(Ω)
(Ω)
Min (Ω)
Max (Ω)
400
360
354.3
365.7
4k
3.6 k
3.543 k
3.657 k
40 k
36  k
35.43 k
36.57 k
400 k
360  k
354.3  k
365.7  k
4 M1
3.6 M
3.543 M
3.657 M
40 M
36  M
35.43 M
36.57 M
Move connection from resistance standard no. 1 to resistance standard no. 2.
1
14


Privacy Statement