TB 9-6625-159-24
Table 31. Model 2608A Temperature.
Performance Checks
Cool/heat source
Test instrument
Output
Indication limits
(C)
Range
Min (C)
Max (C)
-20
TEMP
-25.0
-15.0
0
TEMP
-5.0
+5.0
50
TEMP
45.0
55.0
100
TEMP
95.0
105.0
150
TEMP
140.0
160.0
Table 32 Model 380911F Temperature.
Performance Checks
Cool/heat source
Test instrument
Output
Indication limits
Range
(F)
Min (F)
Max (F)
-4
TEMP
-4.084
-3.916
100
TEMP
97.6
102.4
500
TEMP
489.6
510.4
1000
TEMP
976.0
1024
Table 33. Model 380973 Temperature.
Performance Checks
Cool/heat source
Test instrument
Output
Indication limits
Function/Range
(C)
Min (C)
Max (C)
C
-10
-13.2
-6.8
C
100
95.0
105.0
C
300
291.0
309.0
C
400
384.0
416.0
15. Capacitance
a. Performance Check
(1) Connect capacitance standard to TI capacitance inputs observing polarity.
(2) Set TI function switch to capacitance.
(3) Set TI and capacitance standard to settings listed in table 34 for model 380973.
TI will indicate within limits specified.
b. Adjustments. No adjustments can be made.
Table 34. Model 380973 Capacitance.
Performance Checks
Capacitance standard
Test instrument
Output
Indication limits
Function/Range
(F)
Min (F)
Max (F)
1000
F
965
1035
17